Uni Bremen Home MARUM - Sediment Geochemistry - Leobener Str - D-28359 Bremen - Germany
Marine Geochemistry - Laboratory Methods

inlab methods main

onboard methods main


Access element methods

Ni - Nickel

sample type

marine porewater
digested sediment 1g/L
groundwater
typical concentration
µmol/L
ppb
1 µmol/L
50 ppb

µmol/L
ppb

Method1 ICP-OES
wavelength 231.604 nm -> wavelength table
method used
no
yes
no
dilution
 
-
 
sample size
 
> 2 mL
 
detction limit
 
20 ppb
 
standard solutions
 
Single element standard in nitric acid 1%
 
Method 2 GF-AAS
wavelength 231.604 nm
method used
yes
yes
yes
dilution
1:10
1:1000
-
sample size
50 µL
500 µL
500 µL
detction limit
2 ppb
1ppb
1 ppb
standard solutions
Single element standard in nitric acid 1%
Single element standard in nitric acid 1%
Single element standard in nitric acid 1%
Method 3 ICP-MS
Mass 60 -> Mass Table
method used
yes
yes
yes
dilution
1:100
1:10
1:10
sample size
> 1 mL
> 1 mL
> 1mL
detction limit
10 ppb
1 ppb
1 ppb
standard solutions
Multielement standard in nitric acid 1% M4, M6, Kraft ME
Multielement standard in nitric acid 1% M4, M6, Kraft ME
Multielement standard in nitric acid 1% M4, M6, Kraft ME
tips + tricks  

page created by M.Kölling & T.Feseker